228
1100 Series Nano Pump Service Manual
5
Introduction to the Nano Pump
The Main Power Supply Assembly
The main power supply comprises a closed assembly (no onsite repair
possibility).
The power supply provides all DC voltages used in the module except for the
voltages supplied by the lamp power supply to the deuterium and tungsten
lamps in the detectors. The line voltage can vary in a range from 100 240 volts
AC ± 10 % and needs no manual setting.
To disconnect the instrument from line, unplug the power cord. The power
supply still uses some power, even if the power switch on the front panel is
turned off.
Figure 49
Main Power Supply (MPS) Block diagram
primary
circuits
+24 V
supply
+5 V
switching
regulator
linear
voltage
regulators
+12 V
supply
floating
sense
MPS
Line
voltage
+36 V / 2.5 A (*)
+24 V / 4.5 A (*)
+5 V / 3 A
+15 V /.3 A
-15 V /.3 A
(*) total power consumption on +36 V and +24 V must not exceed 107 watts.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......