1100 Series Nano Pump Service Manual
153
Repairing the Pump
3
Assembling the Main Cover
Frequency
If cover is broken
Tools required
None
Parts required
Cover kit G1312-68703 (includes base, top, left and right)
N O T E
The cover kit contains all parts, but it is not assembled.
WA R N I N G
In case you insert the left or right side in the opposite position, you may not be able
to remove the side from the top part.
1 Place the top part on the bench and insert the left
and right side into the top part.
2 Replace the cover.
3 Replace the nano pump into the stack and reconnect the cables and capillaries.
4 Turn on the nano pump.