1100 Series Nano Pump Service Manual
3
Repairing the Pump
2
Disconnect the waste tube. Beware of leaking solvents due to hydrostatic
pressure.
3
Using the 14 mm wrench unscrew the purge valve and remove it from the
purge valve holder.
4
Remove the plastic cap with the gold seal from the purge valve.
5
Using a pair of tweezers or a toothpick to remove the frit.
6
Place a new frit into the purge valve with the orientation of the frit as
shown above.
7
Replace the cap with the gold seal.
8
Insert the purge valve into the purge valve holder and locate the pump
outlet and the waste outlet as shown below.
9
Tighten the purge valve, reconnect outlet capillary and waste tubing.
Figure 19
Purge Valve Parts
Valve body
Gold seal
PTFE frit
Plastic cap
N O T E
Before reinstallation always check the gold seal. A deformed seal should be exchanged.
N O T E
When fused silica capillaries are used, use the 4 mm wrench together with the torque tool
to avoid overtightening.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......