Description
The test is designed to verify the performance of the EMPV. The test must
always be done when the EMPV valve is exchanged. The test should also be
done if column flow stability problems occurs (micro mode only).
The EMPV test is not a substitute for the leak test or pressure test. The leak
and pressure tests should also be done when leaks within the pump heads
might be the problem.
Step 1
The test starts with a short flushing sequence and a cleaning procedure for the
EMPV.
Step 2
Afterwards, low and high pressure is controlled by the EMPV and the
appropriate current is monitored.
Step 3
Finally, a linear pressure ramp is performed.
Running the test
1
Select the ChemStation’s Diagnosis screen; from the nano pump tests
selection box, select “Micro Mode Pressure Test”.
2
Start the test and follow the online instructions
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......