Troubleshooting and Test Functions
2
Valve Failed
Valve 0 Failed:
valve A1
Valve 1 Failed
: valve A2
Valve 2 Failed
: valve B2
Valve 3 Failed
: valve B1
One of the solvent selection valves in the nano pump failed to switch correctly.
The processor monitors the valve voltage before and after each switching
cycle. If the voltages are outside expected limits, the error message is
generated.
Probable Causes
•
Solvent selection valve disconnected.
•
Connection cable (inside instrument) not connected.
•
Connection cable (inside instrument) defective.
•
Solvent selection valve defective.
Suggested Actions
✔
Ensure the solvent selection valve is connected correctly.
✔
Ensure the connection cable is connected correctly.
✔
Exchange the connection cable.
✔
Exchange the solvent selection valve.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......