1100 Series Nano Pump Service Manual
41
Troubleshooting and Test Functions
2
Pressure Below Lower Limit
The system pressure has fallen below the lower pressure limit.
Probable Causes
•
Lower pressure limit set too high.
•
Air bubbles in the mobile phase.
•
Leak.
•
Defective damper.
•
Defective CSM board.
Suggested Actions
✔
Ensure the lower pressure limit is set to a value suitable for the analysis.
✔
Ensure solvents are degassed. Purge the nano pump.
✔
Ensure solvent inlet filters are not blocked.
✔
Inspect the pump head, capillaries and fittings for signs of a leak.
✔
Purge the nano pump. Run a pressure test to determine whether the seals
or other pump components are defective.
✔
Exchange the damper.
✔
Exchange the CSM board.