1100 Series Nano Pump Service Manual
33
Troubleshooting and Test Functions
2
Leak Sensor Short
The leak sensor in the nano pump has failed (short circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current increases above the upper limit,
the error message is generated.
Probable Causes
•
Defective leak sensor.
•
Leak sensor incorrectly routed, being pinched by a metal component.
Suggested Actions
✔
Exchange the leak sensor.