21
Agilent 1100 Series Nano Pump
Service Manual
Agilent Technologies
2
Troubleshooting and Test Functions
This chapter describes the pump’s built in troubleshooting and test functions:
Status Indicators
The nano pump is provided with two status indicators which indicate the
operational state (prerun, run, and error states) of the nano pump. The status
indicators provide a quick visual check of the operation of the nano pump
(
Error Messages
In the event of an electronic, mechanical or hydraulic failure, the nano pump
generates an error message in the user interface. The following pages describe
the meaning of the error messages. For each message, a short description of
the failure, a list of probable causes of the problem, and a list of suggested
actions to fix the problem are provided (
Micro Pressure Test
This is a fast test to verify the tightness of a micro system, where the pump is
operating in the micro mode and no manual purge valve is installed. The pump
is operating in the pressure control mode at 380 bar for several minutes. The
remaining flow in the column flow path between the EMPV and the blank nut
is measured. (
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......