134
1100 Series Nano Pump Service Manual
3
Repairing the Pump
Exchanging the Nano Pump Main Board (NPM Board)
1
Turn off the pump, disconnect all cables and remove the pump from the
stack.
2
Remove the top covers and foam, see “Removing the Top Cover and Foam”
on page 131.
3
Disconnect all connectors from the main board.
Frequency
Board defective
Tools required
Wrench 14 mm
Wrench 7 mm
Wrench 5 mm
Parts required
NPM board, G2226-66530, exchange part number G2226-69530
Connector
Description
Connector Description
S1
Configuration switch
J1
GPIB
P11
AIV - A
J2
CAN connector
P15
Leak sensor
J3
CAN connector
P16
Damper
J4
RS232
P17
Fan
J5
Remote
P18
Interface board
J6
Analog output
P21
Encoder B
P2
Power supply
P20
AIV-B
P8
Motor B
P22
Solvent Selection Valve
P12
Motor A
P9
EMPV
P14
Encoder A
P19
Flow sensor
N O T E
When removing connectors, counterhold on connector J3 with one hand.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......