1100 Series Nano Pump Service Manual
115
Repairing the Pump
3
Exchanging the Purge Valve Frit or the Complete Manual Purge Valve
Exchanging the purge valve frit or the complete purge valve
1
Using a 1/4 inch or 4 mm wrench disconnect the pump outlet capillary at
the purge valve.
Frequency
Frit
-
when plunger seals are exchanged or when contaminated or blocked
(pressure drop of > 10 bar across the frit at a flow rate of 2500
µ
l/min of H
2
0
with purge valve opened)
Purge valve - if internally leaking
Tools required
Wrench 1/4 inch
Wrench 4 mm (depending on pump outlet capillary)
Wrench 14 mm
Pair of tweezers or toothpick
Parts required
PTFE Frit (pack of 5) 01018-22707
Purge valve G1311-60009
Figure 18
Exchanging the Purge Valve Frit
Purge valve
Pump outlet
capillary
Purge valve holder
Waste capillary
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......