1100 Series Nano Pump Service Manual
3
Repairing the Pump
2
Before inserting the new valve cartridge clean the area in the actuator
assembly. Fill a syringe with alcohol and flush the cartridge area
thoroughly.
3
Insert a new valve cartridge into the actuator assembly. Make sure the valve
cartridge is fully inserted into the actuator assembly.
Replacing the Active Inlet Valve
1
Insert the new valve into the pump head. Using the 14 mm wrench turn the
nut until hand tight.
2
Position the valve that the solvent inlet tube connection points towards the
front.
3
Using the 14 mm wrench tighten the nut by turning the valve in its final
position (not more than a quarter turn). Do not overtighten the valve. The
solvent inlet tube connection should point to the right corner of the pump
head.
4
Reconnect the inlet tube and the active inlet valve cable to the connector at
the Z-panel.
5
After an exchange of the valve cartridge it may take several mL of pumping
with the solvent used in the current application, before the flow stabilizes
at A%-ripple as low as it used to be, when the system was still working
properly.
N O T E
Make sure you are in normal mode.
Summary of Contents for 1100 Series
Page 1: ...s1 Agilent 1100 Series Nano Pump Service Manual ...
Page 4: ...iv 1100 Series Nano Pump Service Manual ...
Page 12: ...xii 1100 Series Nano Pump Service Manual Agilent Technologies on Internet 269 Index 271 ...
Page 32: ...20 1100 Series Nano Pump Service Manual 1 Installing the Pump ...
Page 114: ...102 1100 Series Nano Pump Service Manual 2 Troubleshooting and Test Functions ...
Page 204: ...192 1100 Series Nano Pump Service Manual 4 Parts and Materials ...
Page 242: ...230 1100 Series Nano Pump Service Manual 5 Introduction to the Nano Pump ...
Page 265: ...1100 Series Nano Pump Service Manual 253 Control Module Screens for the Nano Pump 6 ...
Page 272: ...260 1100 Series Nano Pump Service Manual 7 Specifications ...
Page 282: ...270 1100 Series Nano Pump Service Manual A Safety Information ...
Page 287: ......