1100 Series FD Reference Manual
2
First Steps with the Fluorescence Detector
Example: Optimization for Multiple Compounds
Using PNAs as a sample, this example uses the described scanning functions.
Setting the Chromatographic Conditions
1
This example uses the following chromatographic conditions (the detector
settings are shown in
Table 12
Chromatographic Conditions
Mobile phases
A = water = 50%
B = Acetonitrile = 50%
Column
Vydac-C18-PNA, 250 mm x 2.1 mm i.d. with 5 µm
particles
Sample
PAH 0.5 ng
Flow rate
0.4 ml/min
Compressibility A (water)
46
Compressibility B (Acetonitrile)
115
Stroke A and B
auto
Time Table
at 0 minutes %B=50
at 3 minutes %B=60
at 14.5 minutes %B=90
at 22.5 minutes %B=95
Stop time
26 minutes
Post time
8 minutes
Injection volume
1 µl
Oven temperature (1100)
30°C
FLD PMT Gain
PMT = 15
FLD Response time
4 seconds