Capturing Data
94
Agilent 1000B Series Oscilloscopes User’s Guide
4
Continue pressing the the
Coupling
softkey or turn the
entry knob to
select one of these trigger coupling settings:
To set the trigger high-frequency reject coupling
Trigger high- frequency reject coupling (100 kHz cutoff) is used to filter
high frequency signal components from the trigger path when they
interfere with achieving stable triggers.
To set the trigger high- frequency reject coupling:
1
Press
[Menu]
.
2
In the Trigger menu, press
Set Up
.
3
In the Set Up menu, press
HF Reject
to toggle between “ON” and “OFF”.
DC
Sets the trigger coupling to DC.
AC
Sets the trigger coupling to AC — use for waveforms
greater than 50 Hz.
LF Reject
Sets the trigger coupling to low frequency reject (10 kHz
cutoff).