Displaying Data
2
Agilent 1000B Series Oscilloscopes User’s Guide
51
To use a digital filter
You can apply a digital filter to the sampled waveform data.
1
If the channel’s menu is not currently displayed, press the channel key
(
[1]
,
[2]
,
[3]
, or
[4]
).
2
In the Channel menu, press
Digital Filter
.
3
In the Filter menu, press
Filter Type
, and continue pressing the
Filter Type
softkey or turn the
entry knob to select between:
4
Depending on the type of filter selected, press
Upper Limit
and/or
Lower Limit
, and turn the
entry knob to adjust the limit.
The horizontal scale control sets the maximum value for the upper and
lower limits.
Digital filters are not available when:
•
The horizontal scale is 20 ns/div or lower.
•
The horizontal scale is 50 ms/div or higher.
To change the Volts/Div control sensitivity
When you need to adjust the amplitude/div setting in smaller increments,
you can change the sensitivity ofthe vertical scale control.
1
If the channel’s menu is not currently displayed, press the channel key
(
[1]
,
[2]
,
[3]
, or
[4]
).
2
In the Channel menu, press
Volts/Div
to toggle between:
LPF (Low Pass Filter).
HPF (High Pass Filter).
BPF (Band Pass Filter).
BRF (Band Reject Filter).
Summary of Contents for 1000 Series
Page 1: ...s1 Agilent 1000B Series Oscilloscopes User s Guide ...
Page 12: ...12 Agilent 1000B Series Oscilloscopes User s Guide Contents Safety Symbols 150 Index 151 ...
Page 16: ...16 Agilent 1000B Series Oscilloscopes User s Guide Tables ...
Page 36: ...1 Getting Started 36 Agilent 1000B Series Oscilloscopes User s Guide ...
Page 64: ...2 Displaying Data 64 Agilent 1000B Series Oscilloscopes User s Guide ...
Page 98: ...3 Capturing Data 98 Agilent 1000B Series Oscilloscopes User s Guide ...
Page 116: ...4 Making Measurements 116 Agilent 1000B Series Oscilloscopes User s Guide ...