Identifying problems
Please refer to the details of following symptoms or error messages. If the
problem persists, turn off the detector and consult your sales representative or
local dealer.
WARNING:
Improper changes, additions, maintenance or repair of the
system can lead to personal injury, electrical shock and damage
to the equipment. Safety is only guaranteed when changes,
additions, maintenance or repairs are carried out by an Agfa
certified field service engineer. A non certified engineer
performing a modification or service intervention on a medical
device, acts on his own responsibility and makes the warranty
void.
Symptom
Cause
Remedy
The detector will not
turn on.
The battery is
not attached.
Attach the battery.
The battery pack
is not charged.
Fully charge the battery pack.
The battery pack
is broken.
Replace the battery pack.
The status indicator of
the System Control Unit
does not light up.
The power cord
is unplugged
from the AC out-
let.
Connect the plug to the AC
outlet firmly. If it still does not
work, replace the System Con-
trol Unit.
The status indicator of
the System Control Unit
does not light up in
green.
A hardware error
has occurred.
Turn off the System Control
Unit and turn it on again. If it
still does not work, replace the
System Control Unit.
The detector status dis-
play shows an error
number (e.g.
"ERR_01"
).
An error has oc-
curred.
Turn off the DR Detector and
turn it on again. Follow the in-
structions for starting up the
DR detector.
If the error persists call Serv-
ice.
A fully charged battery is
consumed quickly.
The battery ca-
pacity decreases.
The DR Detector battery can
deteriorate because of its char-
acteristics and structure. For
120
| XD 14, XD
⁺
14 | Problem solving