5.4.3.2 Sharpness
Adjustment
(1) Check the DICOM conformance statement whether the modality sends the images
in page mode or image mode.
(2) Select the required smooth factor at the modality to define the sharpness of the
image. See
.
(3) When finished, print test images from the modality.
(4) Check the image quality together with the customer
(5) In case the image quality is not satisfying continue as described in the flowchart.
See Figure 15.
536401XK.CDR
sharp
too sharp
less sharp
299
199
101
200
148
143
245
235
251
138
Cubic High Res
Cubic Bell
Interpolation Smooth Factors (DICOM values)
299
200
251
245
235
Limits
Legend
Default
Useable values
Figure 18
NOTE:
Use
Cubic Bell
interpolation, when the modality sends the images in Page
Mode (images already composed for the film).
Use
Cubic High Res
interpolation, when the modality sends the images in
Image Mode (printer composes the different images for the film).
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......