DD+DIS033.07E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 1 / 25
06-2008 Type
5367/100
Agfa Company Confidential
Host Profile Adjustments for “small clear border” at Mammo printing:
DICOM Host Profile Parameter
"Mammo CWR Rotation" =
yes
"Image Orientation" =
r180
Rotate 180° before
Film tint depends on print direction:
Film 1 has slightly different color tint
if viewed from the side
Film overlap approx. 2 to 4 mm only
hanging at the lightbox
R
L
Lightbox
Print
Direction
1
2
L
R
Film 1
Film 2
-
+
DICOM Host Profile Parameter
"Mammo CWR Rotation" = no (default)
"Image Orientation" = r0 (default)
Same film tint for both films
Film overlap approx. 4 mm:
Fog of blank frame of film 1 adds in film 2
536401gq.cdr
L
R
Lightbox
Without Chest Wall Rotation
Print
Direction
R
L
Film 1
Film 2
-
+
536701gq.cdr
Figure 17