DD+DIS033.07E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 1 / 19
06-2008 Type
5367/100
Agfa Company Confidential
5.2
Logical printers Set-up
(1) Check, whether the printer requires the "logical printer setup", i.e. to setup more
than one AE title for the printer. See flowchart below:
Okay. Logical
Printer Setup not
required.
Okay. Logical Printer
Setup not required.
Perform adjustments
at modality.
Create two logical printers.
Proceeding see this section.
No
Yes
536401mm.cdr
No
Yes
Adjustment
at modality
possible?
Logical Printer
Setup Decision
Flowchart
Check whether modality supports
more than one application (e.g. Mammo + CR).
Check whether images of
different applications can
be adapted to required settings.
Example. DR: Dmax 3.2; LUT: Kanamori
Mammo: Dmax: 3.8, LUT: linear
Modality supports
> 1 application ?