DD+DIS033.07E
Controls, Connections and Setup Procedure
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 1 / 16
06-2008 Type
5367/100
Agfa Company Confidential
NOTE:
It is also possible to enter the site data in
the "site data tool" on the service PC (see
MEDNET, GSO Library:
Hardcopy => Dry Imagers => DRYSTAR
AXYS => Freeware) and upload the so
created ZIP file via key-operator or service
menu "9 Installation - 3 Install site data".
Figure 10
4
Film Calibration
The purpose of this section is to show the tasks that are required for film calibration.
(1)
Load the films the customer will
use in the upper and lower film
tray. Refer to note below.
(2)
Select in key-operator menu
'6 Calibration - 1 Film Calibration'.
(3)
Select 'Start Calibration':
The calibration film as shown in
X
Figure 9
X
is printed and the
measured Dmax is shown.
See also note next page.
INPUT TRAY:
FILM CALIBRATION OK
Dmax = 3.20 (XR310)
Default densitomete
536701aa.cdr
Measured Dmax
(4)
Repeat film calibration for second
input tray.
In order to be useable this film should be
absolutely dust free within the marked region!
CDM measurement area
536536ga.cdr
In ord
er t
o be us
eab
le t
h
is
f
ilm
s
hou
ld be ab
s
ol
u
te
ly
dus
t
fr
e
e w
it
hi
n
t
he m
ar
k
e
d re
gi
on!
Figure 11
NOTE:
The films delivered with the printer may only be used for test purposes.
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......