![AGFA DRYSTAR AXYS Service Manual Download Page 283](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651283.webp)
DD+DIS033.07E
Repair and Service
Adjustments and Calibrations
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
DRYSTAR AXYS
Chapter 3.6 / 31
06-2008 Type
5367/100
Agfa Company Confidential
Service - Calibration
Film calibration
1.
Internal densitometer calibration
2.
Thermal print head calibration
3.
Horizontal alignment
4.
Smart profile calibration
5.
Compliance film calibration
6.
Go back
Step Target values Measured values Step Target values Measured values
00
0.16
01
0.33
02
0.49
03
0.66
04
0.83
05
0.99
06
1.16
07
1.33
08
1.50
09
1.66
10
1.83
11
2.00
12
2.16
13
2.33
14
2.50
15
2.66
16
2.83
17
3.00
Service - Calibration
Compliance film calibration
Start calibration
Go back
Reset values
0.16
0.33
0.49
0.66
0.83
0.99
1.16
1.33
1.50
1.66
1.83
2.00
2.16
2.33
2.50
2.66
2.83
3.00
Service - Calibration
Compliance film calibration
Select densitometer:
Macbeth TR924
Gretag
X-Rite 301
X-Rite 310
Start calibration
Go back
Reset values
5367_SIB04_02.CDR
Service - Calibration
Dmin:
0.20
Dmax:
3.68
Reflected Ambient Light
1
View Box Illumination
4000
Number of steps:
18 (EUREF)
Start calibration
Go back
Reset values
Service - Calibration
Compliance film calibration
1.
Upper tray
2.
Lower tray
Go back
Service - Calibration
Compliance film calibration
Actions which must have taken place with the same tray and with the same film pack:
1. First film calibration.
2. Profile calibration.
3. Second film calibration.
4. Remove film pack in other tray(s).
5. Printed DICOM-image.
6. Measurement of the densities.
Proceed
Go back
Service - Calibration
Compliance film calibration
1.
Reset compliance film calibration for upper tray (8x10)
2.
Reset compliance film calibration for lower tray (14x17)
3.
Proceed without resetting
Go back
Compliance film calibration
1
2
3
4
5
6
7
Compliance film calibration
Figure 18
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......