3.7.3
ADVP as Communication Recorder
ADVP as Communication recorder is recording the communication between the
modality and the printer, and allows to modify every single DICOM attribute as well as
resending the recorded image with the changed attribute(s).
In general, following configurations are possible:
a) Service PC is part of the network: The modality sends to the service PC, the
service PC transmits the DICOM image and control data on the fly to the
printer. Modifying the DICOM data can be done without changing the
configuration.
b) Service PC is only temporary integrated in the network to emulate the printer:
The modality sends to the service PC, ADVP on the service PC transmits the
DICOM image and control data to a DICOM Print SCP on the service PC (e.g.
ADVP as Print Emulator). Re-sending the DICOM data to the printer then
requires re-configuration of ADVP.
In this chapter we describe the possibility a) Service PC is part of the network.
Summary of Contents for DRYSTAR AXYS
Page 7: ......
Page 98: ......
Page 106: ......
Page 250: ......
Page 290: ......
Page 332: ......
Page 334: ......
Page 352: ......
Page 372: ......
Page 378: ......
Page 409: ......
Page 440: ......
Page 455: ...DRYSTAR AXYS Type 5367 100 1st Edition r Orde no I 0 DD D S 33 07E ...
Page 456: ......