![AGFA DRYSTAR AXYS Service Manual Download Page 156](http://html.mh-extra.com/html/agfa/drystar-axys/drystar-axys_service-manual_2866651156.webp)
DD+DIS033.07E
Repair and Service
Tools and Auxiliary Means
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
3.7.1
ADVP as DICOM print SCU
ADVP as DICOM print SCU can be used to check the DICOM interface of the printer.
In this case ADVP sends a test image to the printer.
(1) Start
ADVP
(2) Select 'Basic Grayscale' in the
'Start up configuration' screen.
(3) Select 'Test - Print Film'
(4) Select 'SMPTE' (or any other
test film).
(5) In the 'Configuration' window fill
in Called (Printer) AE Title and
'Printer IP address'.
(6) Select 'OK'. Now the desired
test film is sent to the printer.
Figure 51: ADVP as Print SCU
3.7.2
ADVP as DICOM print SCP (Print Emulator)
ADVP as DICOM print SCP can be used to check the DICOM communication sent
from the modality. ADVP in this case emulates the printer. The communication can be
evaluated on site or by a product specialist in the GSC or RSC.
Edition 1, Revision 0
DRYSTAR AXYS
Chapter 3.2 / 50
10-2007 Type
5367/100
Agfa Company Confidential
Summary of Contents for DRYSTAR AXYS