Controls, Connections and Set Up Procedures
DD+DIS150.05E
Printed copies of this document are not controlled.
The reader should always refer to MedNet to ensure that they are working with the latest copy of the document.
Chapter 1 / 24
Drystar 5302
Edition 1, Revision 0
536401GM.CDR
255
Input
Kana
mor
i
200
100
75
Kanamori like value
Kanamori curve and
Kanamori_like 100 = Kanamori
Kanamori_like 200 = linear
Kanamori_like 220 = logarithmic
Possible Look up table settings
Dmax
220
0
Dmin
Figure 17
536601GL.CDR
5.9
0.1
1.1
0.9
0.7
Limits
Legend
Default
Reasonable values
sharp
too sharp
less sharp
5.9
4.9
-4.9
0.1
-0.2
-0.7
0.9
0.7
1.1
-1.2
Cubic High Res
Cubic Bell
Interpolation Smooth Factors (Kernel values)
Figure 18
For details regarding image adjustments refer to the Hardcopy Application
Manual DD+DIS078.05E.
Each of the image adjustment parameters can be set for a specific modality.
Changing a parameter value will change the value of the respective
parameters in the 'DICOM Host profile’ for that specific modality.
Agfa Company Confidential
(Typ
5366/100)
2005-09-09