DD+DIS150.05E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
9
Software Release 2.0.0_C2
Type of new Software*
Software Upgrade
Release Date
07/2009
Printer Serial Numbers
≥
SN 8923 plus 8634, 8662, 8679, 8691,
8694, 8847,8848, 8854, 8856, 8857, 8858,
8865, 8871, 8873, 8875, 8880, 8886, 8893,
8894, 8896, 8897, 8903, 8908, 8911, 8912,
8913, 8914, 8915, 8916, 8917, 8918, 8919,
8920, 8921
Availability
MedNet, GSO Library
Prerequisites
None
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
On MedNet, GSO Library always the latest available software release can be
downloaded.
9.1
New Features
Software 2.0.0_C2 contains following new features:
Software supports both printers, DRYSTAR 5302 and DRYSTAR AXYS.
Support of 14 bit contrast resolution (instead of 12 bit)
Improved RF-tag locking mechanism (identical to RF-tag locking mechanism of
DRYSTAR 550X software 6.0)
Support of “Compliance film calibration” which may be required to pass the
Afssaps / EUREF* acceptance test.
For DRYSTAR AXYS this feature was available already as of software 1.6.3_C1,
however only for Mammo printing. Now it is available also for non-mammography
applications for DRYSTAR 5302 and DRYSTAR AXYS.
*
Agence française de sécurité sanitaire des produits de santé
=
European Reference Organisation for Quality Assured Breast Screening and
Diagnostic
Services
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.8 / 17
03-2011 Type