Symptom 2
When the upper or lower RF tag (RF = Radio Frequency) is Causeive (Error HC301003
/ HC 302003) the display names the wrong defective RF tag (RF1 instead of RF2 and
vice verse).
Also in the service manual DD+DIS150.05E, section 4, edition 1 revision 0, Reference
Diagrams, the RF-tags are mixed.
Correct:
RF1 = lower tray
RF2 = upper tray
Cause
Mismatch of correct naming of RF tag in translation file for LCD display, service manual
and software
8.2
New Feature
Currently the RF-Tags of the film packages
have their information stored in a non-ISO
format.
To be prepared for future changes of the RF
tags (new format as defined by ISO) software
≥
1.9.2_C1 already accepts ISO RF tags.