DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
8
Software Release 1.9.2_C1
Type of new Software*
Software Upgrade
Release Date
11/2008
Printer Serial Numbers
≥
SN 7069, plus 6827, 6867, 6887, 6902,
6917, 7067, 7052 and 7057
Availability
MedNet, GSO Library
Prerequisites
None
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
On MedNet, GSO Library always the latest available software release can be
downloaded.
8.1
Solved Software Bugs
Software 1.9.2_C1 solves following software bugs:
Symptom 1
Sporadically the printer stops boot-up with message “Error System Failure”:
When pressing the ON/OFF switch the printer cannot be switched off.
Access via network is not possible (to reboot remotely).
After unplugging and reconnecting the mains cable the printer usually starts up
normally.
Cause
CPU Driver problem
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.8 / 15
03-2011 Type
5366/100
Agfa Company Confidential