DD+DIS150.05E
Repair and Service
Software Releases and Patches
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
7
Software Release 1.9.1_C1
Type of new Software*
Software Upgrade
Release Date
04/2008
Printer Serial Numbers
≥
SN 5723 plus 5660, 5665, 5667, 5669,
5672, 5676, 5677, 5679, 5683, 5684, 5717,
5719, 5721
Availability
MedNet, GSO Library
Prerequisites
None
*NOTE:
A software upgrade is a functional extension. An update (software patch) is an
internal change designed to correct bugs, fix defects, or otherwise minor
improvements to performance without changing the software's basic design,
structure or functionality.
On MedNet, GSO Library always the latest available software release can be
downloaded.
7.1
Solved Software Bugs
Software 1.9.1_C2 solves following software bugs:
Symptom 1
Sporadic input film jams for films from the upper tray, mainly 14” films:
Error message at the local display:
Film jam! Open upper input tray and clear film path
Error message in the infocounter: HC314003
Cause
The force for driving the film towards the thermal head is sporadically too low.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 3.8 / 13
03-2011 Type