DD+DIS148.05E
Installation Planning
1
2
3
DOCUMENT CONTROL NOTE:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
8
Electrical connections
Electrical installations have to be done in accordance
with national regulations. Examples:
VDE
(Germany) Electrical installations in the
installation room must be in compliance with the
regulations IEC 364, VDE 0100, and VDE 0107.
Installation guidelines
UL
(US) Electrical installations in the installation
room must be in compliance with the regulations:
“National Electrical Code" (NEC) (NFPA70).
IMPORTANT:
A ground fault interrupter (GFI) (IN = 30 mA) must be installed at a suitable point in
the circuit (in compliance with VDE 664).
VDE
Double earthed pin outlet in compliance with
DIN 49441 and with CEE 7 standard cover V II.
Mains connection in the
installation room
UL
Earthed contact outlet for the NEMA 5-15 P plug
Euro-
Version
:
CEE7/7 - H05VV - F3G - 3 x 2.5 mm² with
IEC C13 plug (female); cable length 3 m
Mains cable
(part of the delivery)
US/CAN-
Version:
NEMA 5-15P - SJT, 3 x AWG 18 with
IEC C13 plug (female); Cable length 3 m
CEE7/7
IEC C13
IEC C13
NEMA 5-15P
Europe
US
5366_chap11_001.cdr
Figure 9
Network cable
(part of the delivery)
UTP; AWG24; Length: 5 m
Main switch
Upon machine installation, it must be ensured that either
the mains connector or an all-pole circuit breaker for the
installation on site is located close to the machine and
easily accessible.
Edition 1, Revision 2
DRYSTAR 5302
Chapter 11 / 21
12-2007 5366/100
Agfa Company Confidential