90
2920B EN 20031206
$
If you select “CubicBell”, the following screen will appear, allowing you to
adjust the smoothing factor:
Enter the smoothing factor value by means of the arrow keys. Refer to
on page
. Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
Changing the maximum density (MaxD)
1
Perform steps 1 to 3 of the
‘Changing Image quality settings for a modality’
procedure on page
to select the modality and choose ‘Select other’.
The ‘Select for Modality’ screen is displayed:
2
In the ‘Select for Modality’ screen, press the Down key twice, followed by the
Confirm key to select ‘MaxD’.
The ‘MaxD (maximum density)’ screen is displayed:
All maximum densities are selectable starting from 2.80 to 3.40 in steps of 0.05
units. The activated setting is marked with a ‘*’.
"
The maximum density will be limited to the maximum density of the media.
3
Press the up/down key to select the desired maximum density setting for this
modality.
4
Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
CS
Enter smoothing
parameter:
advised value = 0.35
range [0.1-10]
enter absolute value
y.yy
CS
Select for
Modality16char:
1 Look-up table
2 Interpolation
3 MaxD
CS
Select MaxD for
Modality16char:
2.80
2.85
...
* 3.10
3.15
...
3.40
Summary of Contents for Drystar 5300
Page 1: ...Drystar 5300 Reference manual POWER ...
Page 160: ...160 2920B EN 20031206 ...
Page 168: ...168 2920B EN 20031206 ...
Page 189: ...Equipment information sheet Appendix A ...
Page 195: ...Drystar media density response data Appendix B ...
Page 198: ...198 2920B EN 20031206 ...
Page 199: ...Quality Control Charts Appendix C ...
Page 205: ...Index Appendix D ...
Page 210: ...Printed in Belgium Published by Agfa Gevaert N V B 2640 Mortsel Belgium 2920B EN 20031206 ...