53
2920B EN 20031206
Viewing general image quality information
1
Perform steps 1 and 2 of the
‘Accessing the Show settings menu’
procedure,
on page
2
On the Show settings menu, press the Down key three times, followed by the
Confirm key to select ‘Image quality’.
The following screen will appear:
3
Press the Confirm key to select ‘General’.
The automatic film calibration settings are displayed:
The following information is displayed on the automatic film calibration settings:
•
whether the automatic film calibration is set (‘ON’, ‘OFF’),
•
when ON, the automatic film calibration frequency (e.g. every ‘05’ film packs),
•
the maximum density result value (e.g. ‘3.10’) of the last successful film
calibration.
4
Press the Confirm key to return to the Image quality menu.
5
In the Image quality menu, press the Escape key to return to the Show
settings menu.
SH
1 General
2 Def. densitometer
3 Modality16charac
4 Modality16charac
5 etc...
Film Calib.: ON
After 05 packs
MaxD result: 3.10
SH
Summary of Contents for Drystar 5300
Page 1: ...Drystar 5300 Reference manual POWER ...
Page 160: ...160 2920B EN 20031206 ...
Page 168: ...168 2920B EN 20031206 ...
Page 189: ...Equipment information sheet Appendix A ...
Page 195: ...Drystar media density response data Appendix B ...
Page 198: ...198 2920B EN 20031206 ...
Page 199: ...Quality Control Charts Appendix C ...
Page 205: ...Index Appendix D ...
Page 210: ...Printed in Belgium Published by Agfa Gevaert N V B 2640 Mortsel Belgium 2920B EN 20031206 ...