Precautions on EMC
Medical electrical equipment needs special precautions regarding EMC and
needs to be installed and put into service according to the EMC information
provided in the manual.
Portable and mobile RF communications equipment can affect medical
electrical equipment.
WARNING:
The DR Detector should not be used adjacent to or stacked with
other equipment and if adjacent or stacked use is necessary, the
DR Detector should be observed to verify normal operation in
the configuration in which it will be used.
WARNING:
Avoid to place the DR Detector too close to life supporting
devices. Keep a minimal distance of 26 cm between the DR
Detector and life supporting devices.
CAUTION:
Using cables and accessories not mentioned in this manual or
spare parts not ordered from Agfa, may cause a higher emission
of electromagnetic phenomena and/or may rise the
susceptibility against it.
DR 10s | Remarks for HF-emission and immunity |