
Calibration Frequency
The DR Detector must be recalibrated at regular intervals. Messages will be
displayed on the NX workstation when recalibration is required.
Perform calibration when exposure conditions have changed significantly and
after a major change to the hardware or software configuration of the X-ray
system.
A DR detector used for veterinary applications has a calibration interval of 12
months, regardless of the model.
Table 1: Calibration interval
DX-D 10C, DX-D 10G, DX-D 20C, DX-D 20G
3 months
DR 18M, DR 24M
DX-D 40C, DX-D 40G, DX-D 45C, DX-D 45G,
DX-D 60C, DX-D 60G
XD 10, XD
⁺
10, XD 14, XD
⁺
14, XD 17, XD
⁺
17
6 months
DR 10e C, DR 14e C, DR 17e C
DR 10s, DR 14s
DX-D 25C, DX-D 30C, DX-D 35C
Fixed DR Detector (Pixium RF4343 FL)
12 months
Fixed DR Detector (4343R, Pixium RAD 4343) 3 months or 12 months,
depending on the DR De-
tector model.
A timely reminder is dis-
played on the NX worksta-
tion.
DR 14e G, DR 17e G
no calibration required
Table 2: Using the DR Detector on multiple NX workstations
DR 10s, DR 14s
DX-D 10C, DX-D 10G, DX-D
20C, DX-D 20G
DX-D 25C
DX-D 30C, DX-D 35C
Each NX workstation has its own set of cali-
bration data. The regular calibration must be
performed on each of the workstations.
DR Detector Calibration | Calibrating the DR Detector |
9
0134I EN 20210924 1204