Generator controls are disabled
Problem
The calibration of a detector for dynamic imaging cannot be
continued because the generator controls in the software con-
sole are suddenly disabled.
Cause
A communication problem with the generator has occurred.
Work-
around
1.
Abort the calibration procedure.
2.
Restart the calibration.
3.
The initial uniformity check has to be repeated, but the
modes for which calibration was already performed, can be
skipped.
42
| DR Detector Calibration | Problem solving
0134I EN 20210924 1204