Black border and cropping on NX
The black border and cropping on NX can fail on the multiple exposures.
• No, not all or only partial diagnostic images are visible.
• The area between the single intra oral images and the unexposed image
area appears white.
To view the images:
1.
Turn off the
Black border
functionality at NX.
2.
Draw a manual collimation area for each intra oral image on the scan.
3.
Apply the
Collimation
functionality at NX.