77
2241A EN 20040210
General procedure in case of malfunction
In exceptional situations the CR 75.0 display provides comprehensive
information concerning errors and ways of correcting them. The Digitizer
status changes from ‘READY’ to one of the following:
Message
Status
indicator
Meaning
Action
Warning
Red
flashing
Further operation is
possible without impairing
the image quality.
Follow the instructions on
the display. The warning
disappears as soon as
the problem has been
solved.
Locked
Red
flashing
The Digitizer no longer
takes cassettes from the
input buffer. You can
solve this problem without
resetting the Digitizer.
Follow the instructions on
the display.
Error
Constant
red
This status normally
requires service or key
operator intervention.
Follow the instructions on
the display.
Never reset the Digitizer to solve a cassette or image plate jam nor to
solve communication problems with the image processing station.