DD+DIS155.11E
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Evaluation:
If the scanned flatfield does
not
show one of the artifacts, no further action
is required.
If the scanned flatfield shows one of the artifacts compare the flatfield with the
limit pattern. For detailed instructions refer to chapter 3.3, Troubleshooting
5.5
Performing the Installation Wizard of the Digitizer
(1) Start the browser at the NX
workstation (A).
(2) Enter the digitizer IP
address
192.192.192.192
in the
address window (B).
(3) Enter username and
password obtained in the
digitizer web based training.
Start
User Name:
Password:
OK
Cancel
Authentication Required
192.192.192.192
Start
A
B
C
515101zf.cdr
Figure 66
(4) Follow the installation
wizard.
Do not change the digitizer
network settings and the
destination settings. This is
done later, if required. See
section 4.2.
Digitizer Settings
Digitizer
starts up automatically:
Installation Wizards
Adapt local settings
Site Specific Data
Installation Date Confirmation
Destination Settings
install_poster_17a.cdr
Next
Do not change
Next
Adapt
Next
Confirm installation
Figure 67
Edition 1, Revision 4 CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Chapter 1 / Page 45 of 47
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential