System Integration
The system can be installed "standalone" or can be integrated into the local network:
515100ad.cdr
Digitizer Network Cable
Hospital Network
NX Workstation
Standalone Installation
(Direct network connection)
Integrated Installation
(Part of local network)
NX Workstation
Digitizer
Standalone
Installation:
The digitizer can be connected
directly to the NX workstation. In
this case no other network access
to the NX workstation is possible.
The digitizer network settings are
pre-configured for this setup.
Integrated
Installation:
After network setup of digitizer and
NX workstation the system can be
integrated into the local network.
Figure 6
The following table lists the information which is required if the digitizer is integrated
into the local network.
Parameter:
Default value:
Value:
IP Address:
192.192.192.192
..............................................
Subnet Mask:
255.255.255.0
..............................................
Gateway IP Address*: 192.192.100.1
..............................................
AE Title:
DIG_DEFAULT
..............................................
Hostname: 192.192.192.192
..............................................
* Only required if a network gateway is used on site.