DD+DIS155.11E
Repair and Service
Electrical and mechanical Codes, Fuses, LEDs
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
3
Status Indicator on the User Interface Board
Status
Indicator
515101ag.cdr
The digitizer status is indicated via static
or blinking blue, green and red LEDs.
See table below.
Figure 3
Status
Indicator
Digitizer Status
Meaning
Constant
blue
Ready for erasure cycle
When entering the next image plate it will be erased.
Blue
blinking
Erasure cycle
Image plate gets erased.
Constant
green
Stand-by mode (READY) Ready for scanning.
Green
blinking
Busy
Scan cycle active.
Constant
red
Fatal digitizer error
Service mode active
(e.g. shading
calibration)
Scanning not possible:
Service intervention required or
Service function is called up by the field service engineer which
blocks scanning.
For a list of service functions which block scanning refer to
Service Manual chapter 3.2, Tools and auxiliary Means.
Slow red
blinking
(
1 / sec.)
Boot-up
Digitizer warning or
error
Scanning not possible:
User has to wait until boot-up is finished or
User intervention required to continue.
Fast red
blinking
(
3 / sec.)
No connection to Error
Viewer on PC
Scanning not possible:
Error Viewer not up-to-date or not started up.
Triple
blinking
(
3 / sec. +
1 sec. off)
No network connection
to PC
Scanning not possible:
Ping sent from digitizer to NX workstation with IP address
configured in digitizer failed.
Edition 1, Revision 2
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.4 / Page 7 of 7
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential