DD+DIS155.11E
Repair and Service
Electrical and mechanical Codes, Fuses, LEDs
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
2
Power Board / Power II Board
P3
P1
9
P21
P15
P1
6
L188
L189
P2
2
C189
C267
P5
Q41
R191
L1
#6
F27
P3
S1
S3
ON
S1
1
S3
2
P18
5151_C34_001.cdr
Figure 2: Location of switches on the Power Board / Power II Board
IMPORTANT:
Fuses are
not
displayed on purpose.
Do not
replace any fuse. In case of a broken fuse, replace the board instead.
2.1
Switches on the Power Board / Power II Board
Meaning
S1/1 Not
used.
S1/2 Not
used.
S3
Board reset. Only for R&D purpose.
Edition 1, Revision 2
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.4 / Page 6 of 7
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential