DD+DIS155.11E
Repair and Service
Troubleshooting
DOCUMENT CONTROL NOTE:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
After bootup of the digitizer the network settings are as follows:
IP Address:
192.192.192.192
Subnet Mask:
255.255.255.0
Gateway IP Address:
192.192.192.1
Digitizer AE Title:
DIG_DEFAULT
Digitizer Hostname:
192.192.192.192
NOTE:
After a reset of the network settings it is required to configure the network settings
again. Refer to Service Manual Chapter 1, "Controls, Connections, and Setup
Procedures".
6
Image Faults parallel to Image Plate Run
6.1
Calibration Lines
Symptom
Calibration lines are visible in the clinical
image or a flatfield image.
Figure 3: Example for calibration lines
Cause
Digitizer not calibrated after exchange of optic module or PMT with light collector.
Solution
Perform a shading calibration. Refer to chapter 3.6, Adjustments and Calibrations
Edition 1, Revision 2
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Chapter 3.3 / 13 of 46
03-2013
Type 5151 / 100 / 110 / 200 / 210
Agfa Company Confidential