Exposure Controls
1
2
2
3
1.
The exposure button is used to obtain a static image, to acquire a rapid
sequence or to acquire a tomosynthesis sequence, depending on the mode
that is selected in the software console.
The exposure button is available on the console with the position controls.
The console is located in the operator room and another console can be
located in the examination room.
The exposure button can be pressed and hold down to the first pressure
point to prepare the X-ray tube for exposure.
The exposure button is pressed and hold down fully to make the exposure
or during acquisition of the rapid sequence.
2.
The fluoroscopy pedal is used to acquire a fluo sequence.
The fluoroscopy pedal is available in the operator room as a single pedal
or together with a radiography pedal. Another fluoroscopy pedal can be
located in the examination room together with a radiography pedal.
The fluoroscopy pedal is pressed during acquisition of the fluo sequence.
3.
The radiography pedal is used to obtain a static image or to acquire a rapid
sequence, depending on the mode that is selected in the software console.
The radiography pedal can be available in the examination room together
with a fluoroscopy pedal.
The radiography pedal is pressed to make the exposure or during
acquisition of the rapid sequence.
Figure 12: Exposure controls
28
| DR 800 | Introduction
0392C EN 20210309 1049
Summary of Contents for 6010/200
Page 1: ...DR 800 6010 200 User Manual 0392C EN 20210309 1049 ...
Page 95: ...DR 800 Safety Directions 95 0392C EN 20210309 1049 ...
Page 116: ...Resetting the fluo timer on page 203 116 DR 800 Basic Workflow 0392C EN 20210309 1049 ...
Page 123: ...DR 800 Basic Workflow 123 0392C EN 20210309 1049 ...
Page 217: ...Patient size mAs Extra Large normal mAs 4 DR 800 Software Console 217 0392C EN 20210309 1049 ...