Wireless console with position controls cannot restore
connection
Details
The wireless console is moved outside of the bluetooth communi-
cation range of the radiographic table and the connection is lost.
Afterwards, the wireless console is moved within the bluetooth
commmunication range of the radiographic table again.
The indicator light for the status of the wireless connection keeps
blinking, indicating that the connection is not restored.
Cause
Failure to setup the bluetooth connection.
Brief Sol-
ution
Stop the wireless console using the power button at the foot of
the console and start it again.