Density
These buttons are used to adjust the AEC cut-off dose (and patient entrance
dose accordingly).
Density can be increased and decreased in a range of -4 to +4, representing a
change of -50% to +100% in dose. When disabled, the density range number
appears in black.
Table 32: Dose variation compared to reference dose
(D)
Dose
-4
0.50
-3
0.63
-2
0.75
-1
0.88
0
1 (reference dose)
+1
1.25
+2
1.50
+3
1.75
+4
2.00