Generator Technical Data
Manufactur-
er
Spellman High Voltage Electronics GmbH
Josef-Baumann-Strasse 23
D-44805 Bochum, Germany
Supported
Models
EDITOR HFe 401
EDITOR
HFe 501
EDITOR
HFe 601
EDITOR
HFe 801
Max. Power 40 kW
50 kW
65 kW
80 kW
Power input configurations and maximum input current
Power line
208-240
VAC
Single-phase
Two-phase
(*)
208-240
VAC 50/60
Hz
-
275 A
Power line
400 VAC Y-
source
-
400V 3N~ PE (Y) 50/60 Hz
92 A
113 A
144 A
180 A
Power line
400/480
VAC Delta
source
-
400/480V 3~PE (delta without N) 50/60Hz
The power setting is selected during installation
and printed on the type label.
79 A
97 A
124 A
154 A
Power consumption
Maximum
power
62 kVA
76 kVA
96 kVA
120 kVA
Average
power
(nominal)
20 kVA
24 kVA
35 kVA
44 kVA
Standby
power (> 10
min and de-
pending on
input volt-
age, without
MUSICA Ac-
quisition
worksta-
tion)
250 W
440-490 W
440-540 W
440-560 W
Valory (system with floor mounted column) | Product Information |