Valory technical data
Manufacturer
Agfa NV
Septestraat 27
2640 Mortsel, Belgium
Type
5540/100
Power line
See Generator Technical Data
Permanent filtration
E7254FX X-ray
tube
2.8 mm Al @75kVp
(+ 0.2 mm Al with DAP meter integrated in the collimator)
E7884X and
E7252X X-ray
tube
2.9 mm Al @ 75kVp
(+ 0.2 mm Al with DAP meter integrated in the collimator)
Environmental conditions
Table 34: Environmental conditions for the X-ray system
Environmental Conditions (during storage and transport)
Temperature (ambient)
between -15° and 55° Celsius
Humidity (non condensing)
between 15 and 90 % relative humidity
Atmospheric pressure
between 70 and 106 kPa
Environmental Conditions (during normal operation)
Temperature (ambient)
between 10° and 35° Celsius
Humidity (non condensing)
between 30 and 75 % relative humidity
Atmospheric pressure
between 70 and 106 kPa
Maximum altitude
3000 m
For overall system environmental conditions, the environmental conditions of
the DR Detector should be taken into account. Refer to the related User
Manual for environmental conditions for the DR Detector. When using the DR
Detector inside the bucky, take into account that the temperature inside the
bucky can be up to 8°C higher than the temperature in the X-ray room.
186
| Valory (system with floor mounted column) | Product Information