Loading of the bucky in the table
To load the bucky with a detector:
1
44
2
33
55
6
6
Figure 82: Loading of the bucky in the radiographic table
1.
Open the bucky drawer completely by pulling the front handle.
2.
Open the clamps by pulling the front slider.
3.
Position the detector between the clamps.
CAUTION:
Make sure your fingers are not between the clamping
mechanism and the detector. The clamping mechanism may
hurt your fingers, therefore take special care.
4.
Align the center of the bucky to the center mark on the clamps.
5.
Close the clamps to fix the detector position.
6.
Lock the clamps by turning the lock lever clockwise.
Figure 83: Lock icon
7.
Close the bucky drawer using the front handle.
Make sure the bucky drawer is pushed up to the end to close completely.