Positioning the Radiographic Wall Stand
To adjust the height manually, release the brake for vertical movement by
pressing and holding the button at the side panel of the bucky. The bucky can
be moved up and down using the handgrip.
To stop movement and lock the bucky into position, release the button.
1
1.
Button to release the brake for the manual movement
Figure 77: Positioning controls
DANGER:
Make sure that no persons or objects are within the movement
area of the system where they can collide with moving parts of
the system.
WARNING:
Maintain visual contact with the patient while moving the
equipment close to the patient in order to detect hazardous
situations (e.g. collisions) early and to avoid them.
WARNING:
Be careful not to squeeze your finger or hand. Keep your hands
at the handles while positioning the system.
CAUTION:
The maximum load for the wall stand movement in vertical
direction is 20 kg. The bucky unit may slip downwards when
applying excessive load.
Note
:
Do not move the bucky with excessive force to the end
stop positions.
Related Links
on page 29
on page 190
144
| Valory (system with floor mounted column) | Positioning the Radiographic Wall Stand
0426A EN 20220804 1858