Centering the tube head on the wall stand
To move the X-ray tube head to the center of the DR detector in the bucky of
the radiographic wall stand:
1.
On the X-ray tube head, while pressing the transversal movement button,
move the X-ray tube towards the radiographic wall stand.
Figure 62: Transversal movement button
There is a stop position for centering on the radiographic wall stand.
2.
While pressing the tilting button, tilt the X-ray tube 90° towards the
radiographic wall stand.
Figure 63: Tilting button
3.
While pressing the longitudinal movement button, move the X-ray tube to
the required SID.
Figure 64: Longitudinal movement button
There is a stop position for the default SID.
4.
Press the automatic centering and position tracking button.
Figure 65: Position tracking button
The position tracking icon is displayed.
Figure 66: Position tracking icon
A stop position is enabled, corresponding to the center of the radiographic
wall stand.