– English –
Mounting and commissioning
DIT 10
35
6.6
Zero correction
1.
Close the display unit by pressing the upper part of the housing
onto the lower part of the housing.
The display unit was switched on when the battery was connected.
The display toggles between "Zero" and the current offset of the
pressure sensor (indication in hPa = mbar). The arrows
indicate
that the display unit in calibration mode.
2.
Press the two keys
and
simultaneously to correct the off-
set to the value 0.00.
The
pressure sensor must not be in the tank
when you do
so. In this state, zero calibration can be performed any number
of times.
3.
Press the F key to terminate zero correction.
An arrow pointing to the unit "Litres" is displayed at the bottom
of the display.
6.7
Mounting the pressure sensor
Zero calibration has been performed.
1.
Plug the star onto the sensor; note the position of the ribs at the
star.
2.
Use the spacer to screw the star to the probe.
1
Spacer
2
Stern
3
Sensor
Fig. 5: Pressure sensor assembly
3.
Bend the arms of the star over the spacer as shown in fig. 15.
4.
Push the pressure sensor from the top though the tank connec-
tion thread.