Troubleshooting
Europress
29
9 Troubleshooting
Repairs may only be performed by a specialised company, refer to
chapter 2.4, page 6.
Table 2: Troubleshooting
Problem
Possible reason
Repair
Green LED “Supply
ON” is not on.
No mains voltage
available.
Check mains voltage.
Leak.
Check hoses.
Notify the installation
company.
Test/measure-
ment valve in
position "Test/
Vent".
Set the test valve and
the measurement
valve to position
"Normal Operation".
Red LED “Alarm”
lights up.
Condensate
trap/drying filter
bowl open.
Close condensate
trap and drying filter
bowl.
The yellow LED
"Service" flashes.
Annual mainte-
nance required.
Perform annual main-
tenance, see chap-
ter 8.1, page 27.
Drying beads are
colourless.
Replace drying
beads.
Filter polluted.
-
Replace filter.
Other malfunctions
-
Send the device to
the manufacturer.